| Publication Date |
1995 |
| Personal Author |
Gupta, D.; Wang, L.; Hanssen, L. M.; Hsia, J. J.; Datla, R. U. |
| Page Count |
58 |
| Abstract |
Standard Reference Material (SRM) 1921 is a matte finish polystyrene film and is intended for use in calibrating the wavelength scale of spectrophotometers in the infrared (IR) spectral region from 545.49 cm exp(-1) to 3082.19 cm exp(-1) (18.332 micrometers to 3.2445 micrometers). Thirteen absorption peak positions obtained using a center-of-gravity method are certified. The expanded uncertainty value associated with these peak values are between 0.06 cm exp(-1) and 0.66 cm exp(-1), except for 1.84 at 2850.13 cm exp(-1) and 12.29 at 545.49 cm exp(-1). This publication describes the IR spectrophotometer, instrument calibration, SRM material, measurement procedure, calibration of polystyrene films, and uncertainty determination. |
| Keywords |
|
| Source Agency |
|
| NTIS Subject Category |
|
| Corporate Authors |
National Inst. of Standards and Technology (PL), Gaithersburg, MD. Radiometric Physics Div. |
| Supplemental Notes |
Also available from Supt. of Docs. |
| Document Type |
Technical Report |
| Title Note |
Special pub. |
| NTIS Issue Number |
199517 |