The National Technical Information Service acquires, indexes, abstracts, and archives the largest collection of U.S. government-sponsored technical reports in existence. The NTRL offers online, free and open access to these authenticated government technical reports. Technical reports and documents in its repository may be available online for free either from the issuing federal agency, the U.S. Government Publishing Office’s Federal Digital System website, or through search engines.




Details
Actions:

Evaluation of Gas Phase Titration Technique as Used for Calibration of Nitrogen Dioxide Chemiluminescence Analyzers.


PB242294

Publication Date 1975
Personal Author Ellis, E. C.; Margeson, J. H.
Page Count 49
Abstract A detailed method write-up describing the chemiluminescence procedure for the continuous measurement of nitrogen dioxide in ambient air was developed. Atmospheric concentrations of NO2 are measured indirectly by the chemiluminescence reaction of nitric oxide (NO) with ozone (O3). The NO2 is first thermally reduced to NO before it is reacted with O3. The reliability of measurements made by a continuous or instrumental sampling method is strongly affected by its calibration. In the chemiluminescence procedure, an NO2 analyzer is calibrated by the gas phase titration of NO with O3 that produces NO2 stoichiometrically. The gas phase titration calibration procedure was subjected to a ruggedness test. The results indicate that normal variations in such factors as reaction and mixing chamber volumes, ratio of dilution air flow to flow through the ozone generator, use of difference ozone generators with different dilution air flows, and use of different standard NO levels had little effect on the gas phase titration procedure.
Keywords
  • Nitrogen dioxide
  • Gas analysis
  • Calibration
  • Ozone
  • Chemiluminescence
  • Gases
  • Volumetric analysis
  • Performance evaluation
  • Chemiluminescence method
Source Agency
  • Environmental Protection Agency Office of Research and Development
NTIS Subject Category
  • 99A - Analytical Chemistry
Corporate Authors National Environmental Research Center, Research Triangle Park, N.C. Quality Assurance and Environmental Monitoring Lab.
Document Type Technical Report
Title Note Final rept.
NTIS Issue Number 197517
Evaluation of Gas Phase Titration Technique as Used for Calibration of Nitrogen Dioxide Chemiluminescence Analyzers.
Evaluation of Gas Phase Titration Technique as Used for Calibration of Nitrogen Dioxide Chemiluminescence Analyzers.
PB242294

  • Nitrogen dioxide
  • Gas analysis
  • Calibration
  • Ozone
  • Chemiluminescence
  • Gases
  • Volumetric analysis
  • Performance evaluation
  • Chemiluminescence method
  • Environmental Protection Agency Office of Research and Development
  • 99A - Analytical Chemistry
Loading