National Technical Reports Library - NTRL

National Technical Reports Library

The National Technical Information Service acquires, indexes, abstracts, and archives the largest collection of U.S. government-sponsored technical reports in existence. The NTRL offers online, free and open access to these authenticated government technical reports. Technical reports and documents in its repository may be available online for free either from the issuing federal agency, the U.S. Government Publishing Office’s Federal Digital System website, or through search engines.




Details
Actions:
Download PDFDownload PDF
Download

Lawrence Livermore National Laboratory's Book of Minimum Detectable Activity for Direct Measurement of Internally Desposited Radionuclides in Radiation Workers.


DE2009945587

Publication Date 2008
Personal Author Hickman, D. P.
Page Count 28
Abstract Lawrence Livermore National Laboratory maintains an in vivo measurement program designed to identify and evaluate the activity of radionuclides deposited in the body. Two types of systems are primarily used for the routine monitoring of radiation workers, the lung counting system and the scanning bed whole body counting system. The lung counting system is comprised of two Canberra ACTII detector sets. Each ACTII set contains two planar germanium detectors with carbon composite end windows optimized to measure low energy photon emitting radionuclides. The ACTII detectors are placed on the upper torso over the lungs for the direct measurement of internally deposited radionuclides in the lungs that emit low energy photons. A correction for the thickness of the chest wall is applied to the efficiency. Because the thickness of the chest wall is a key factor in the measurement of low energy photon emitting radionuclides in the lung, the minimum detectable activity is a function of the chest wall thickness. The scanning bed whole body counting system is comprised of a thin air mattress on top of a carbon fiber bed that slowly scans over four high purity germanium detectors. The scanning system is designed to minimize variations in detected activity due to radionuclide distribution in the body. The scanning bed detection system is typically used for the measurement of internally deposited radionuclides that emit photons above 100 to 200 keV. MDAs have been generated for radionuclides that provide energies above 80 keV since the lowest calibration energy for the system is approximately 86 keV.
Keywords
  • Lungs
  • Docimetry
  • Radiation protection
  • Radionuclides
  • Radiation workers
  • Thorax
  • Chest
  • Detection
  • Deposition
  • Lung count minimum detectible activity
Source Agency
  • Technical Information Center Oak Ridge Tennessee
Corporate Authors Lawrence Livermore National Lab., CA.; Department of Energy, Washington, DC.
Supplemental Notes Sponsored by Department of Energy, Washington, DC.
Document Type Technical Report
NTIS Issue Number 200912
Lawrence Livermore National Laboratory's Book of Minimum Detectable Activity for Direct Measurement of Internally Desposited Radionuclides in Radiation Workers.
Lawrence Livermore National Laboratory's Book of Minimum Detectable Activity for Direct Measurement of Internally Desposited Radionuclides in Radiation Workers.
DE2009945587

  • Lungs
  • Docimetry
  • Radiation protection
  • Radionuclides
  • Radiation workers
  • Thorax
  • Chest
  • Detection
  • Deposition
  • Lung count minimum detectible activity
  • Technical Information Center Oak Ridge Tennessee
Loading