The National Technical Information Service acquires, indexes, abstracts, and archives the largest collection of U.S. government-sponsored technical reports in existence. The NTRL offers online, free and open access to these authenticated government technical reports. Technical reports and documents in its repository may be available online for free either from the issuing federal agency, the U.S. Government Publishing Office’s Federal Digital System website, or through search engines.




Details
Actions:

Remote Working Level Monitor. Final Report.


ANL77-XX90

Publication Date 1977
Page Count 167
Abstract The Remote Working Level Monitor (RWLM) is an instrument used to remotely monitor the RN-daughter concentrations and the Working Level (WL). It is an ac powered, microprocessor based instrument which multiplexes two independent detector units to a single central processor unit (CPU). The CPU controls the actuation of the detector units and processes and outputs the data received from these remote detector units. The remote detector units are fully automated and require no manual operation once they are set up. They detect and separate the alpha emitters of RaA and RaC' as well as detecting the beta emitters of RaB and RaC. The resultant pulses from these detected radioisotopes are transmitted to the CPU for processing. The programmed microprocessor performs the mathematical manipulations necessary to output accurate Rn-daughter concentrations and the WL. A special subroutine within the program enables the RWLM to run and output a calibration procedure on command. The data resulting from this request can then be processed in a separate program on most computers capable of BASIC programming. The calibration program results in the derivation of coefficients and beta efficiencies which provides calibrated coefficients and beta efficiencies. (ERA citation 03:033634)
Keywords
  • Radiation monitors
  • Radon
  • Alpha detection
  • Automation
  • Beta detection
  • Calibration
  • Daughter products
  • Design
  • Efficiency
  • Electronic circuits
  • Microprocessors
  • Multiplexers
  • Radiation monitoring
  • Scintillation counters
  • Si semiconductor detectors
  • ERDA/440101
  • ERDA/500300
Source Agency
  • Technical Information Center Oak Ridge Tennessee
NTIS Subject Category
  • 77E - Nuclear Instrumentation
Corporate Authors Argonne National Lab., ILL.; Department of Energy.
Document Type Technical Report
NTIS Issue Number 197819
Contract Number
  • W-31-109-ENG-38
Remote Working Level Monitor. Final Report.
Remote Working Level Monitor. Final Report.
ANL77-XX90

  • Radiation monitors
  • Radon
  • Alpha detection
  • Automation
  • Beta detection
  • Calibration
  • Daughter products
  • Design
  • Efficiency
  • Electronic circuits
  • Microprocessors
  • Multiplexers
  • Radiation monitoring
  • Scintillation counters
  • Si semiconductor detectors
  • ERDA/440101
  • ERDA/500300
  • Technical Information Center Oak Ridge Tennessee